| Model Number |
1443-2511 |
| Medical Device Problem Codes |
Fracture (1260); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reporter states that the surgeon implanted a 2.5 crosstie, part number 1443-2511 lot number 105448.The surgeon implanted it and was going to begin suture and grabbed the toe and the implant broke in half.The surgeon pulled one out and the other end was drilled out.Removal and reimplant was about 4 minutes long.An alternate implant was placed with minimal surgical delay.No non-conformances were identified in the investigation.The root cause is that the crosstie implant is a small diameter implant with potential to fracture if over-stressed.
|
| |
|
Event or Problem Description
|
|
The reporter states that the surgeon implanted a 2.5 crosstie, part number 1443-2511 lot number 105448.The surgeon implanted it and was going to begin suture and grabbed the toe and the implant broke in half.The surgeon pulled one out and the other end was drilled out.Removal and reimplant was about 4 minutes long.An alternate implant was placed with minimal surgical delay.
|
| |
|