| Model Number |
15 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device is not completing the boot up process.As a result, defibrillation therapy would be unavailable, if needed.There was no report of patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned to stryker for evaluation.The cause of the reported issue could not be determined.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker evaluated the customer's device and verified the reported issue.Stryker determined that the cause of the reported issue was due to a failed system pcb assembly.A replacement for this part is no longer available due to part obsolescence.The device was returned to the customer unrepaired.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device is not completing the boot up process.As a result, defibrillation therapy would be unavailable, if needed.There was no report of patient use associated with the reported event.
|
| |
|