MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 HI-WALL E1 LINER 36MM H; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 HI-WALL E1 LINER 36MM H; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a hip revision, the neutral and high wall liners did not seat.There was no soft tissue between the cup and liner and the screw was flush into the cup.The surgery was completed with a dual mobility liner.No adverse consequences were reported.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
Cmp-(b)(4).Report source foreign: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2022-02231.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.Visual examination of the provided pictures identified two g7 liners, no other observation could be made related to this event.Dhr was reviewed and no discrepancies related to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional event information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 HI-WALL E1 LINER 36MM H
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15501885
Report Number0001825034-2022-02230
Device Sequence Number20193446
Product Code PBI
UDI-Device Identifier00880304527133
UDI-Public(01)00880304527133(17)240704(10)6579275
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K121874
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/04/2024
Device Model NumberN/A
Device Catalogue Number010000938
Device Lot Number6579275
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/07/2022
Supplement Date Received by Manufacturer10/25/2022
Initial Report FDA Received Date09/29/2022
Supplement Report FDA Received Date11/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CAT#: 00877503602 BIOLOX HEAD LOT#: 3112567; CAT#: 010000860 G7 LINER LOT#: 6572975; CAT#:110010249 G7 SHELL LOT#: 7046217
Patient SexMale
-
-