Model Number 45-480420-L
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A search for non-conformances associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was not returned to the manufacturer for analysis; therefore, the alleged product issue cannot be confirmed.If the device or additional information is received, microvention will issue a supplemental report.The instructions for use (ifu) identifies premature coil detachment as potential complications associated with use of the device.
 
Event or Problem Description
It was reported that during an embolization procedure, there was resistance when advancing the coil implant in the microcatheter and the coil unintentionally detached inside the microcatheter.The coil was removed from the patient and procedure completed using another coil.There was no report of harm or injury to the patient.
 
Additional Manufacturer Narrative
Summary device evaluation: the investigation of the returned coil system found the pusher kinked at the distal section, and the implant not attached to the pusher.The pusher's heater coil showed no signs of activation using a detachment controller.The implant was not returned for evaluation; however, the investigation of the pusher found the monofilament to be damaged.The monofilament experienced a tensile break based on the profile of the tip, which is consistent with the device experiencing excessive force that exceeded the strength of the monofilament causing the implant to separate from the pusher.The physical evaluation of the device could not identify the conditions or circumstances that led to the damage, but the damage is consistent with the device experiencing forces over specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZUR-18 DETACHABLE HYDRO
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise drive
aliso viejo, CA 92656
7142748000
MDR Report Key15502226
Report Number2032493-2022-00378
Device Sequence Number15392350
Product Code KRD
UDI-Device Identifier00810170017926
UDI-Public(01)00810170017926(11)210719(17)260630(10)2107195KB
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K162524
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2026
Device Model Number45-480420-L
Device Catalogue Number45-480420
Device Lot Number2107195KB
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2022
Supplement Date Received by Manufacturer10/24/2022
Initial Report FDA Received Date09/29/2022
Supplement Report FDA Received Date11/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-