• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE CR RP INSRT SZ 5 8MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE CR RP INSRT SZ 5 8MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT Back to Search Results
Model Number 1516-30-508
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994); Joint Laxity (4526); Unspecified Tissue Injury (4559)
Date of Event 09/14/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Patient underwent a primary attune tkr on the (b)(6) 2022.During the primary surgery the fellow nicked the mcl but elected to continue with the planned implant construct as it was in his opinion a "minor nick" and the knee felt stable and balanced at trial and final implantation.The patient started partial weight bearing during standard post op mobility rehabilitation, and the mcl became more deficient and the cr rp insert dislocated.The patient was brought back today for a full revision to an attune crs.If other, describe: n/a.If other, describe: revision knee.Patient status/ outcome / consequences: yes.Patient consequence description/was there a clinical outcome experienced by the patient (infection, inflammation, etc.)? pain, instability, revision surgery.Was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: yes.If yes, describe: x-rays, revision surgery.Other information: n/a.Is the patient part of a clinical study: unknown.Ip-01532990: device property of: none.Device in possession of: none.Ip-01532991: device property of: none.Device in possession of: none.Ip-01532992: device property of: none.Device in possession of: none.By checking this box i certify that all information that are known/available has been disclosed.If any new information will be made available, the additional information will be submitted through cst.True.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information received indicated that the patient was revised due to infection.No delay reported.Insert dislocated due to mcl deficiency.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CR RP INSRT SZ 5 8MM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15502863
Report Number1818910-2022-19044
Device Sequence Number12720656
Product Code NJL
UDI-Device Identifier10603295047773
UDI-Public10603295047773
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1516-30-508
Device Catalogue Number151630508
Device Lot Number9631059
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2022
Supplement Date Received by Manufacturer10/14/2022
10/20/2022
Initial Report FDA Received Date09/29/2022
Supplement Report FDA Received Date10/19/2022
10/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/09/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE CR FEM RT SZ 5 CEM; ATTUNE RP TIB BASE SZ 5 CEM; CEMENT MANUFACTURER UNKNOWN
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-