SYNTHES GMBH CHUCK WITH KEY-REAM SPEED F/BPL; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
| Catalog Number |
530.732 |
| Medical Device Problem Codes |
Complete Loss of Power (4015); Excessive Heating (4030)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported from russia that during service and evaluation, it was determined that the chuck with key devicecolor band was chipping, was heating up, the k wire settings would not change and the etching was illegible.The device also failed pretests for general condition, markings, check drill chuck function, drill chuck in smallest range, drill chuck in largest range and check for run-out inspection in running mode.It was noted in the service order that the device did not work.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The actual device was returned for evaluation.The device was evaluated and it was determined that the reported condition was not confirmed.Therefore, the assignable root cause was not determined.However, during evaluation, it was determined that the device color band was chipping, was heating up, the k wire settings would not change and the etching was illegible.The device also failed pretests for general condition, markings, check drill chuck function, drill chuck in smallest range, drill chuck in largest range and check for run-out inspection in running mode.The assignable root cause could not be determined.Udi: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|