Model Number LOGIC TIB INSERT IMPL CRC, SZ 3, 15MM
Medical Device Problem Codes Mechanical Problem (1384); Unstable (1667); Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Loss of Range of Motion (2032); Scar Tissue (2060); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Balance Problems (4401); Joint Laxity (4526); Implant Pain (4561); Swelling/ Edema (4577)
Date of Event 05/18/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No device return anticipated due to this being a legal case.Concomitants: 4123716, 200-02-32 - three peg patella 32mm; 4184945, 02-012-45-3030 - lgc tibial fit tray cem sz 3f / 3t; 4186241, 02-010-03-0330 - logic cr femoral cem, right, sz 3.Recall: z-0021-2022.Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Event or Problem Description
It was reported via a legal notification that a female patient, initial right total knee replacement surgery, (b)(6) 2016, began to suffer from symptoms associated with the defects of exactech knee system as alleged in this complaint, including pain and lack of mobility.As a result of these symptoms, plaintiff underwent revision surgery on (b)(6) 2022 approximately 6 years 4 months post the initial procedure.No further information known.
 
Additional Manufacturer Narrative
H6: based on review of all available information, there is no evidence to suggest that the reported event is related to any design issues.The cause of the pain and subsequent revision cannot be conclusively determined; however, it is most likely related to the patient¿s underlying condition.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, instability, loss of range of motion, femoral loosening, and tibial loosening or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC TIB INSERT IMPL CRC, SZ 3, 15MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
3523771140
MDR Report Key15503037
Report Number1038671-2022-01196
Device Sequence Number20772158
Product Code JWH
UDI-Device Identifier10885862230430
UDI-Public10885862230430
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K123342
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/25/2021
Device Model NumberLOGIC TIB INSERT IMPL CRC, SZ 3, 15MM
Device Catalogue Number02-012-51-3015
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2022
Supplement Date Received by Manufacturer10/11/2022
02/10/2025
Initial Report FDA Received Date09/29/2022
Supplement Report FDA Received Date11/07/2022
02/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/30/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-