• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) JOT DX¿ ICM; IMPLANTABLE CARDIAC MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) JOT DX¿ ICM; IMPLANTABLE CARDIAC MONITOR Back to Search Results
Model Number DM4500
Device Problem Premature Discharge of Battery (1057)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/08/2022
Event Type  malfunction  
Event Description
It was reported that the patient presented for an implant procedure.The insertable cardiac monitor (icm) to be implanted exhibited premature battery depletion.The icm was not implanted.Patient was stable.
 
Manufacturer Narrative
The reported event of premature discharge of battery was not confirmed.The device was received in normal working conditions with battery voltage above elective replacement indicator (eri).Analysis performed, indicated normal device functionality.Longevity assessment was performed, and the device was in the normal range of operation with appropriate remaining longevity.A device history record (dhr) review was performed, and all required manufacturing processes and inspection steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.
 
Event Description
New information notes that no replacement product was implanted as the procedure was cancelled for reason unrelated to product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOT DX¿ ICM
Type of Device
IMPLANTABLE CARDIAC MONITOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15503729
MDR Text Key305277231
Report Number2017865-2022-39600
Device Sequence Number1
Product Code MXC
UDI-Device Identifier05415067040046
UDI-Public05415067040046
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 11/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/14/2022
Device Model NumberDM4500
Device Catalogue NumberDM4500
Device Lot NumberS000083045
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/12/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/21/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-