MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ALTRUA 60; IMPLANTABLE PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ALTRUA 60; IMPLANTABLE PACEMAKER Back to Search Results
Model Number S602
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/16/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that battery status indicators were different than expected based on information provided to boston scientific.Factors influencing the estimated longevity remaining calculation include pacing rate, amplitude, pulse-width and lead impedance.Any changes in these factors will impact the battery consumption calculation and therefore the remaining longevity estimate.Please refer to the describe event or problem for more information regarding the specific circumstances of this event.Please note this pacemaker's battery was designed to estimate remaining longevity based on the pacing capacitor charge time, which progressively increases as the battery depletes.These pacemakers were not designed with an ability to calculate remaining longevity for every possible current drain, which may cause replacement indicators to trigger earlier than what was previously estimated; however, it does not negatively impact the use, operation, safety, or performance of the device.
 
Event or Problem Description
It was reported that the magnet rate of this pacemaker dropped to ninety beats per minute and then back to a hundred beats per minute after several months.A request was made to have data from this device analyzed.However, there was no memory data for the reset counters is available.The memory data returned for the device appears to be incomplete and missing key memory locations required to perform an accurate analysis of battery performance.This device has been implanted for approximately 11.2 years and minimum expected longevity is approximately 7 years.Since the device has far exceeded minimum expected longevity calculations, it was concluded that elective replacement time (ert) declaration may occur at any time.Further, boston scientific technical services (ts) consultant discussed results with health care professional and sounds like will opt to continue the 7 week checks.To date, this device remains in service.No adverse patient effects reported.
 
Event or Problem Description
It was reported that the magnet rate of this pacemaker dropped to ninety beats per minute and then back to a hundred beats per minute after several months.A request was made to have data from this device analyzed.However, there was no memory data for the reset counters is available.The memory data returned for the device appears to be incomplete and missing key memory locations required to perform an accurate analysis of battery performance.This device has been implanted for approximately 11.2 years and minimum expected longevity is approximately 7 years.Since the device has far exceeded minimum expected longevity calculations, it was concluded that elective replacement time (ert) declaration may occur at any time.Further, boston scientific technical services (ts) consultant discussed results with health care professional and sounds like will opt to continue the 7 week checks.To date, this device remains in service.No adverse patient effects reported.Additional information received indicates that the device was explanted.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that battery status indicators were different than expected based on information provided to boston scientific.Factors influencing the estimated longevity remaining calculation include pacing rate, amplitude, pulse-width and lead impedance.Any changes in these factors will impact the battery consumption calculation and therefore the remaining longevity estimate.Please refer to the describe event or problem for more information regarding the specific circumstances of this event.Please note this pacemaker's battery was designed to estimate remaining longevity based on the pacing capacitor charge time, which progressively increases as the battery depletes.These pacemakers were not designed with an ability to calculate remaining longevity for every possible current drain, which may cause replacement indicators to trigger earlier than what was previously estimated; however, it does not negatively impact the use, operation, safety, or performance of the device.This product has not been returned for analysis.This report will be updated should additional information become available or if analysis is completed.
 
Event or Problem Description
It was reported that the magnet rate of this pacemaker dropped to ninety beats per minute and then back to a hundred beats per minute after several months.A request was made to have data from this device analyzed.However, there was no memory data for the reset counters is available.The memory data returned for the device appears to be incomplete and missing key memory locations required to perform an accurate analysis of battery performance.This device has been implanted for approximately 11.2 years and minimum expected longevity is approximately 7 years.Since the device has far exceeded minimum expected longevity calculations, it was concluded that elective replacement time (ert) declaration may occur at any time.Further, boston scientific technical services (ts) consultant discussed results with health care professional and sounds like will opt to continue the 7 week checks.To date, this device remains in service.No adverse patient effects reported.Additional information received indicates that the device was explanted.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that battery status indicators were different than expected based on information provided to boston scientific.Factors influencing the estimated longevity remaining calculation include pacing rate, amplitude, pulse-width and lead impedance.Any changes in these factors will impact the battery consumption calculation and therefore the remaining longevity estimate.Please refer to the describe event or problem for more information regarding the specific circumstances of this event.Please note this pacemaker's battery was designed to estimate remaining longevity based on the pacing capacitor charge time, which progressively increases as the battery depletes.These pacemakers were not designed with an ability to calculate remaining longevity for every possible current drain, which may cause replacement indicators to trigger earlier than what was previously estimated; however, it does not negatively impact the use, operation, safety, or performance of the device.This product has not been returned for analysis.This report will be updated should additional information become available or if analysis is completed.The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.The device was then exposed to simulated heart load conditions, pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTRUA 60
Common Device Name
IMPLANTABLE PACEMAKER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key15503867
Report Number2124215-2022-38536
Device Sequence Number12701416
Product Code LWP
UDI-Device Identifier00802526484476
UDI-Public00802526484476
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
N970003/S096
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/06/2011
Device Model NumberS602
Device Catalogue NumberS602
Device Lot Number586723
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/09/2022
Supplement Date Received by Manufacturer11/01/2023
04/29/2024
Initial Report FDA Received Date09/29/2022
Supplement Report FDA Received Date11/09/2023
05/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/07/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age71 YR
Patient SexMale
-
-