• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH 1.9F ARGYLE SINGLE LUMEN PICC; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH 1.9F ARGYLE SINGLE LUMEN PICC; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number 43303
Medical Device Problem Codes Fluid/Blood Leak (1250); Separation Problem (4043)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.Device evaluated by mfr: device evaluation comment - the customer stated that the device will not be returned for evaluation but did not provide rationale.
 
Event or Problem Description
Customer reports a failure of the picc catheter.Per additional information received on 09/19/22, a breakage and leaking occurred just below the butterfly where the catheter is joined.Chlorohexadine was used to clean the area prior to insertion, it was then secured with secureport iv adhesive & sorbavion dressing.The product was inserted on (b)(6) 2022 on the right upper extremity basilic and removed on (b)(6) 2022.The device was replaced with another 1.9fr picc line.The patient is doing well.
 
Additional Manufacturer Narrative
The reported lot number was invalid, therefore a manufacturing device history review (dhr) or product/process changes review for the involved lot number could not be performed.However, all dhrs are reviewed to meet regulatory requirements prior to product release.The actual sample involved in the reported incident was not returned for evaluation.No additional information, photographs or videos were received.Consequently, it was not possible to evaluate it as part of a comprehensive failure investigation and confirm the reported condition.Based on current manufacturing process, no manufacturing or design issues were identified.As such, the root cause could not be determined based on the available information.It must be noted that in-process controls such as personnel training, incoming quality acceptance testing for raw material, 100% in process visual inspection and visual acceptance sampling are performed in the plant are in place to prevent nonconforming product from leaving the manufacturing operations.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.9F ARGYLE SINGLE LUMEN PICC
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
edificio b20 calle #2 zona fra
alajuela
CS  
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key15504034
Report Number3009211636-2022-00796
Device Sequence Number5224502
Product Code LJS
UDI-Device Identifier10884527004645
UDI-Public10884527004645
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number43303
Device Catalogue Number43303
Device Lot Number11343966
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/16/2022
Supplement Date Received by Manufacturer09/16/2022
Initial Report FDA Received Date09/29/2022
Supplement Report FDA Received Date10/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-