| Model Number |
9004051 |
| Medical Device Problem Codes |
Device Alarm System (1012); Defective Component (2292)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A product sample was received and is awaiting evaluation, investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the capnography monitor gave a co2 sensor alarm.Information pointed to the co2 bench on the transducer having gone bad and needing to be replaced.There has been no report of patient involvement or no observable clinical symptoms or a change in symptoms identified in the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: the device was returned for evaluation.The keypad was detached from the front bezel.Functional tests indicated that the co2 bench lost its factory calibration memory data.The root cause is undetermined.Factory calibration was performed to clear the sensor error.(h6c, h6d, h6e).Additional event information received indicating no patient involvement and date of event was received.A manufacturing device history record review was not performed, due to the age of the device.
|
| |
|
Event or Problem Description
|
|
Additional information was received via email on 02-oct-2022.Event date was updated.No patient was involved.
|
| |
|
Search Alerts/Recalls
|