| Model Number |
4517 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
This cardiac resynchronization therapy pacemaker (crt-p) and left ventricular (lv) lead triggered a lead safety switch (lss) to unipolar due to a high out-of-range pace impedance measurement of 2,012 ohms.The vector was reprogrammed successfully and the lss for the lv lead was programmed off.This crt-p and lv lead remain in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
This cardiac resynchronization therapy pacemaker (crt-p) and left ventricular (lv) lead triggered a lead safety switch (lss) to unipolar due to a high out-of-range pace impedance measurement of 2,012 ohms.The vector was reprogrammed successfully and the lss for the lv lead was programmed off.This crt-p and lv lead remain in service.No adverse patient effects were reported.Additional information received reported that the cardiac resynchronization therapy pacemaker (crt-p) was explanted and replaced due to normal battery depletion (nbd).In addition, a left ventricular (lv) adapter was placed.This left ventricular (lv) lead remains in service.It was noted that the lv pace impedance measurement with the new crt-p was 1142 ohms.No additional adverse patient effects were reported.
|
| |
|
Search Alerts/Recalls
|