SMITH & NEPHEW, INC. JRNY II BCS FEM COCR LT SZ 4; PROST, KNEE, PATFEMTIB, SEMI-CNSTRND, CMNTD, POL/MET/POL Back to Search Results
Model Number 74024924
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 08/29/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, during a tka, the impactor was inadvertently placed upside down on the jrny ii bcs femoral cocr lt sz 4.The problem was noticed once the implant was placed on the femur with cement, and the impactor did not dislodge.Surgeon attempted to dislodge it with a mallet, which then caused a fracture of the lateral condyle.The fracture was repaired with three (3) screws and remainder of procedure completed.The same impactor was used and surgery was prolonged for 15-20 min.Fracture caused patient prolonged recovery with only partial weight bearing, including unexpected rehab hospital care.
 
Additional Manufacturer Narrative
Initial mdr report.Internal reference: (b)(4).
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that based on the documentation provided, a user/procedural error contributed to the reported event.No clinical factors have been identified as a potential root cause.The patient impact included the reported bone fracture with subsequent 3-screw reduction, surgical delay, ¿prolonged recovery with only partial weight bearing, including unexpected rehab hospital care, and is still ongoing¿.Further patient impact cannot be determined.No further medical assessment can be rendered at this time.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for knee systems revealed in indications, contraindications, and adverse effects that the preoperative planning and meticulous surgical technique are essential to achieve optimum results.Considerations of component placement are critical to minimize a variety of postoperative complications.Warnings and precaution section states that during curing of cement, care should be taken to prevent movement of the implant components.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors that could contribute to the reported event include surgical technique used or user/procedural variance.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II BCS FEM COCR LT SZ 4
Common Device Name
PROST, KNEE, PATFEMTIB, SEMI-CNSTRND, CMNTD, POL/MET/POL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15505394
Report Number1020279-2022-04215
Device Sequence Number5224553
Product Code JWH
UDI-Device Identifier00885556611425
UDI-Public00885556611425
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K111711
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number74024924
Device Catalogue Number74024924
Device Lot NumberA1902993
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/29/2022
Supplement Date Received by Manufacturer11/10/2022
Initial Report FDA Received Date09/29/2022
Supplement Report FDA Received Date11/14/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/08/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age76 YR
Patient SexFemale
-
-