MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAFLEX MACHINES; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAFLEX MACHINES; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 955052
Device Problem Device Alarm System (1012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/02/2022
Event Type  malfunction  
Event Description
It was reported that during continuous renal replacement therapy with a prismaflex control unit and a prismaflex set, an alarm condition related to self-test failure 1 was reportedly generated.The machine and circuit were changed out after losing last circuit due to same issue.The treatment was ended without the extracorporeal blood being returned to the patient both times.It was further reported the pressure sensors were cleaned; however, they did not dry the sensor.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
The device was not returned and the lot number is unknown; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX MACHINES
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - LUND MONITORS
magistratsvagen 16
lund skane lan SE-22 643
SW   SE-22643
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key15505443
MDR Text Key306415019
Report Number9616026-2022-00096
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 10/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number955052
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/02/2022
Initial Date FDA Received09/29/2022
Supplement Dates Manufacturer Received10/21/2022
Supplement Dates FDA Received10/27/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
PRISMAFLEX SET
-
-