The article, "aortic migration of amplatzer piccolo ductal occluder", was reviewed.The research article presents a case study on a 4 month old former 23 week premature baby, 610grams underwent device closure of a large patent ductus arteriosus in the cardiac catheterization lab with an amplatzer piccolo device at 2.4kg.He patent ductus arteriosus measured 2.5 mm at its narrowest dimension and 8.1 mm long and was closed using a 5-2 amplatzer piccolo occluder, based on the proctor¿s recommendation.Angiography during the procedure showed a device which sits largely intraductually.Intra-procedural transthoracic echocardiography showed no evidence of obstruction and hence the device was deployed in this position.Following the release of device part of the aortic disc protruded into the aorta while most of the device is still in the ductus.No attempt was made at retrieving or repositioning the device at this point since on echocardiogram there appeared to be no obstruction.Post procedure echocardiogram also showed no evidence of pulmonary or aortic obstruction 6 hours, 24 hours, and 1-week post-procedure.A follow-up echocardiogram performed 4 weeks post-procedure showed interval migration of the device into the aorta causing obstruction.The patient was taken back to the cardiac catheterization lab where significant obstruction with a peak gradient of 50 mmhg in the aortic isthmus was confirmed.Multiple attempts were made to snare the device from the pulmonary end but this was not successful since the pin and the full pulmonary disc were in the ductus.The aortic pin was snared and attempts were made to recapture the device and reposition it but the device could not be moved and instead the entire aorta moved with the device.The patient was transferred to our hospital for surgical removal of the device in light of the severe obstruction of the aorta.An echocardiogram performed at our institution confirmed severe obstruction of the aorta.The patient underwent removal of ductal device occluder, closure of the patent ductus arteriosus and repair of the left pulmonary artery and aorta at 18 weeks of life.The decision was made to perform the operation on cardiopulmonary bypass.The patent ductus arteriosus was dissected along its length and the device could be partially seen and felt from outside.Circulatory arrest was then initiated, and the patent ductus arteriosus was incised opened longitudinally to expose the pulmonary disc.The ductal tissue around the device was freed gently.The body of the piccolo device and the aortic end of the disc which had migrated well into the aortic lumen were gently freed of all adhesions.The device was then explanted in its entirety by traction on the pulmonary disc.Completion transesophageal echocardiogram demonstrated good ventricular function with a normal doppler pattern in the descending thoracic aorta.Following confirmation of hemostasis, the chest was closed.The patient did well in the post-operative period and follow-up echocardiograms continue to demonstrate normal ventricular function with no aortic obstruction.The article concluded that appropriate selection of device size is crucial for good outcomes.Aortic migration and obstruction are potentially serious complications of this device.Close and extended monitoring is required to diagnose this complication.[the primary and correspondence author is prashant k.Minocha, md, department of pediatric cardiology, nyu langone health, new york, ny 10016, email address: prashant.Minocha@nyulangone.Org].
|