The investigation determined that higher than expected total b-hcg ii (bhcg) results were obtained from a non-vitros biorad quality control (qc) fluid using vitros immunodiagnostics products total bhcg ii reagent lot 3441 in combination with a vitros 5600 integrated system.The assignable cause for the higher than expected vitros bhcg qc fluid results is suboptimal calibrations.The parameters for all the calibration events that occurred from (b)(6) 2022 were atypical when compared to database values.The cause of the suboptimal calibrations is unknown.After restoring the calibration event from 25 (b)(6) 2022, vitros bhcg lot 3441 qc results returned to expectation.Based on acceptable historical quality control results, a vitros bhcg lot 3441 performance issue is not a likely contributor to the event.Additionally, although there was no diagnostic precision testing run on the vitros 5600 system, an instrument issue is not a likely contributor to the event as acceptable results were obtained before and after the atypical calibration events with no actions performed on the analyzer.Continual tracking and trending of complaints has not identified any signals that would indicate a potential systematic issue with vitros bhcg ii reagent lot 3441.Email address for contact office above is (b)(6).
|
A customer contacted the ortho clinical diagnostics (ortho) technical solution center (tsc) to report higher than expected b-hcg ii (bhcg) results were obtained from a non-vitros biorad quality control (qc) fluid using vitros immunodiagnostics products total bhcg ii reagent lot 3441 in combination with a vitros 5600 integrated system.Biorad lot 85210 level 1 vitros bhcg results of 9.8, 10.4, 14.8, 10.0 and 9.2 miu/ml versus an expected result of 6.12 miu/ml biased results of the magnitude and direction observed may lead to inappropriate physician action if they were to occur undetected on patient samples.The higher than expected vitros bhcg results were from a non-patient fluid.The customer did not give any indication that patient results had been affected.There is no allegation of patient harm as a result of this event.This report corresponds to ortho clinical diagnostics inc (ortho) complaint number (b)(4).
|