| Model Number |
VTICM5_12.6 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No similar complaint was reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated the surgeon implanted a 12.6mm vticm5_12.6 implantable collamer lens, -11.00/+2.5/109 (sphere/cylinder/axis), into the patients eye on (b)(6) 2022.The lens was removed on (b)(6) 2022 due to lens rotation.The lens was replaced with a different lens.Additional information has been requested but none has been forthcoming.
|
| |
|
Additional Manufacturer Narrative
|
|
B5: the lens was implanted into the patients right eye (od).The lens was replaced with a longer lens and the problem was resolved.Cause of the event was unknown.Claim # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3: device evaluation: the lens was returned in liquid, in a specimen cup.Visual inspection found a haptic torn.Dimensional verification was performed and the lens was found to be in specification.Claim # (b)(4).
|
| |
|