| Catalog Number |
MZ9265 |
| Medical Device Problem Code |
Defective Component (2292)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported while using bd microbore bi-fuse extension set, 2 iv co the tubing separated from the luer.There was no report of patient impact.The following information was provided by the initial reporter: bi-tubing extension broke away from conjunction site.
|
| |
|
Additional Manufacturer Narrative
|
|
A complaint of tubing separating at the conjunction site was received from the customer.A photo was provided for investigation.Separation is seen at the part where the set splits into two separate tubing lines.The customer complaint was confirmed.A device history record review for model mz9265 lot number 21105858 was performed.The search showed that a total of 7,203 units in 1 lot number were built on (b)(6) 2021.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.A root cause could not be determined as a physical sample was not received.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|