| Model Number |
Coronavirus |
| Medical Device Problem Codes |
False Positive Result (1227); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Customer reporting 4 false positive sars results.Customer states the results were confirmed negative by another molecular method.It is noted the customer started experiencing issues after a 3rd party piece of equipment was replaced and is also having the negative qc control result positive.Report 3 of 4.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: customer data was reviewed and analysis did not conclude a product performance issue.Analysis revealed a higher fam signal attributed to post-run analysis of the data.Quidel provided modified programing for the third-party equipment to mitigate the issue moving forward.Root cause: no product problem found source: email/phone.
|
| |
|