Model Number CD1211-36
Medical Device Problem Code Inappropriate or Unexpected Reset (2959)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/12/2022
Type of Reportable Event Serious Injury
Event or Problem Description
During an in clinic follow up, it was noted the device was in back up mode.Technical support was contacted and suspected battery was low and eri (elective replacement indicator) had been reached.It was noted a device firmware download was carried out.The device was explanted and replaced to resolve the event.The patient was stable.
 
Additional Manufacturer Narrative
The reported event of backup operation was confirmed.Analysis of device image found the device went into backup mode due to a known mechanism that can be encountered when too many high voltage charges are performed in too short a period of time.After the device was restored from backup mode, functional testing was performed.Telemetry, impedance, sensing, and pacing functions of the device were tested and found to be normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CURRENT PLUS VR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15506794
Report Number2017865-2022-39812
Device Sequence Number5069832
Product Code NVZ
UDI-Device Identifier05414734503686
UDI-Public05414734503686
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2016
Device Model NumberCD1211-36
Device Catalogue NumberCD1211-36
Device Lot NumberS000000242
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/12/2022
Supplement Date Received by Manufacturer12/02/2022
Initial Report FDA Received Date09/29/2022
Supplement Report FDA Received Date12/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/09/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-