This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
As reported, during an unspecified urological procedure, the laser came through the sides of a cook single-use holmium laser fiber.The light was escaping about half way down the fiber.Another fiber was used to complete the procedure.The device did not make contact with the patient, so no harm was caused.
|
|
Event summary as reported, during an unspecified urological procedure, the laser came through the sides of a cook single-use holmium laser fiber.The light was escaping about halfway down the fiber.Another fiber was used to complete the procedure.The device did not make contact with the patient, so no harm was caused.Investigation - evaluation reviews of the complaint history, device history record, manufacturing instructions, and quality control procedures, as well as a visual inspection and functional test were conducted during the investigation.The device was returned in open packaging for investigation.The laser fiber was connected to testing laser unit.The laser beam appeared to shine through the fiber shaft approximately 27.2cm from distal end.The laser fiber was kinked at the point where the light is escaping from.A document-based investigation evaluation was performed.No related non-conformances were recorded, and no other complaints have been reported for this product lot.The device history record review provides objective evidence that the device was manufactured to specification.There is no evidence of nonconforming devices from the complaint lot in house or in the field.A review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was functionally inspected by quality control and no notable gaps in production or processing controls were noted.There is no indication that a design or process related failure mode contributed to the reported event.Sufficient inspection activities are in place to identify this failure mode prior to distribution.Based on the available information, cook has concluded that it is possible the device was kinked during packaging, shipping/storage, removal from packaging, or the procedure.A definitive cause could not be determined.Cook will continue monitoring of similar complaints and has notified the appropriate personnel of this event.Per a review of risk documentation, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|