As the lot number for the device was provided, a review of the device history record is currently being performed.The device has not been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.Medical device expiry date: 04/2024.
|
|
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.Five photos were reviewed.The first photo shows the label of the device on which the reported product details could be verified.The second photo shows the detached balloon and catheter with exposed inner guide wire lumen.The third photo shows the distal portion of the catheter with exposed inner guide lumen and marker bands.The fourth photo shows the detached balloon, and the fifth photo shows the device stored within a plastic bag.No other anomalies were noted.Based on the submitted photos, the investigation for the reported balloon rupture remains inconclusive as no evidence of rupture could be observed due to the condition of the detached balloon.However, the investigation was able to be confirmed for the reported detachment as the balloon was noted to be completely detached from the catheter shaft and was able to be observed in the submitted photos.A definitive root cause for the reported balloon rupture and balloon detachment could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 04/2024), g3, h6 (method) h11: b5, h6 (result, conclusion) h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|