BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 035; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number U3575102
Medical Device Problem Codes Difficult to Insert (1316); Difficult to Remove (1528); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/09/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an angioplasty procedure in the right common iliac artery, the pta balloon allegedly unable to be advanced and difficult to be removed from the sheath.It was further reported that the device was removed along with the sheath.There was no reported patient injury.
 
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 07/2024).
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one ultraverse 035 pta dilatation catheter loaded in an unknown sheath was returned for evaluation.No other specific anomalies were noted during the visual evaluation.The balloon was noted to be stuck and no functional testing was performed due to the condition of the device.The investigation remains inconclusive for the reported sheath insertion failure into the catheter as the functional testing couldn¿t be performed due to the condition of the device returned for evaluation.However, the investigation was confirmed for the reported sheath removal failure from the catheter as the balloon was noted to be stuck within the catheter and returned for evaluation.A definitive root cause for the reported sheath insertion failure into the catheter and reported sheath removal failure from the catheter could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 07/2024), g3.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during an angioplasty procedure in the right common iliac artery, the pta balloon allegedly unable to be advanced and difficult to be removed from the sheath.It was further reported that the device was removed along with the sheath.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAVERSE 035
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe 85281
Manufacturer (Section G)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford N A
EI   N A
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15507318
Report Number2020394-2022-00773
Device Sequence Number17410801
Product Code LIT
UDI-Device Identifier00801741094576
UDI-Public(01)00801741094576
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K142261
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberU3575102
Device Catalogue NumberU3575102
Device Lot NumberCMFV0304
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/09/2022
Supplement Date Received by Manufacturer11/03/2022
Initial Report FDA Received Date09/29/2022
Supplement Report FDA Received Date11/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-