| Model Number |
FEM12120 |
| Medical Device Problem Code |
Misfire (2532)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history record is currently being performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 02/2025).
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent graft placement procedure, the device allegedly partial deployed.The procedure was completed using another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.Investigation summary: the returned sample analysis could not be performed as the physical device was not returned and no images were provided for evaluation.Based on information available and as no device sample was returned, the investigation is closed with inconclusive result.A definitive root cause could not be determined based upon the available information.Labeling review: in reviewing the relevant labeling, it was found that the instructions for use sufficiently address the potential risks.Regarding the preparation of the device the instructions for use states ¿carefully remove the endovascular system from its packaging and inspect packaging and system for any damage or defects.Do not use if the sterile barrier is compromised', 'flush the stent graft lumen with sterile saline by using a small volume syringe (¿) close the stopcock when flushing is complete and remove the syringe from the luer port'.Regarding the precautions prior and during deployment the instructions for use states 'prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure¿, ¿higher deployment force may be encountered (¿) in tortuous anatomy¿, and 'if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device.¿ the instructions for use states regarding the accessories 'the use of an appropriately sized introducer sheath is recommended; prepare a stiff 0.035" guidewire per its instructions for use and advance the guidewire under fluoroscopy to the target location' the packaging pictograms indicate an introducer size of 10f and a 0.035" guidewire.H10: d4 (expiry date: 02/2025), g3.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent graft placement procedure, the device allegedly partial deployed.The procedure was completed using another device.There was no reported patient injury.
|
| |
|