• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRUMPF MEDIZIN SYSTEME GMBH + CO. KG ILED 7 CEILING TRIO; SURGICAL LIGHT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRUMPF MEDIZIN SYSTEME GMBH + CO. KG ILED 7 CEILING TRIO; SURGICAL LIGHT Back to Search Results
Model Number 4068310
Medical Device Problem Codes Mechanical Problem (1384); Detachment of Device or Device Component (2907); No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/22/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported the plate with vesa monitor adapter and handpiece holder fell during clinical activity which is part of the monitor holder adapter from the company cim.No harm or impact to a surgical procedure was reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Additional Manufacturer Narrative
Inspection of the device is pending.Further investigation should confirm the root cause.Investigation results will be provided within a final report if new information will become available.
 
Additional Manufacturer Narrative
The alleged event was investigated by trumpf medical/hillrom and the supplier of the monitor holder, cim med.During a review of the provided information from our italian distributor (altamed) as well as the returned monitor holder the following conclusion was made: according to the condition of the monitor holder it is not likely that the product was installed and in use.There was no signs of use or wear and parts were still secured by the original packaging.The initial allegation ¿monitor adapter and handpiece holder fell during clinical activity¿ could not be confirmed.The damages found are based on a fall of the equipment from approximately 70 cm, but an event could not have occurred during use or clinical activity.Further questions for the duration of clinical use (which was alleged) were not answered by the distributor/customer.Based on these findings no further actions are necessary.
 
Event or Problem Description
The customer reported the plate with vesa monitor adapter and handpiece holder fell during clinical activity which is part of the monitor holder adapter from the company cim.No harm or impact to a surgical procedure was reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ILED 7 CEILING TRIO
Common Device Name
SURGICAL LIGHT
Manufacturer (Section D)
TRUMPF MEDIZIN SYSTEME GMBH + CO. KG
7-9 carl zeiss strasse
saalfeld 07318
GM  07318
Manufacturer Contact
frances coote
7-9 carl zeiss strasse
saalfeld 07318
GM   07318
MDR Report Key15507924
Report Number9681407-2022-00022
Device Sequence Number14932680
Product Code FSY
UDI-Device Identifier00887761995840
UDI-Public887761995840
Combination Product (Y/N)N
Initial Reporter CountryIT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number4068310
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/22/2022
Supplement Date Received by Manufacturer09/22/2022
Initial Report FDA Received Date09/29/2022
Supplement Report FDA Received Date12/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1