Model Number PS2504-N
Medical Device Problem Code Improper Flow or Infusion (2954)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
Fill volume: 300 ml, flow rate: 5ml/hr, procedure: renal transplant, cathplace: unknown , infusion start time: on (b)(6) 2022; 14:00, infusion stop time: on (b)(6) 2022; 7:00.The lead nurse reported, a pump over-infused, [the infusion] should have lasted around 60 hrs but the ball was empty at around 41 hours post-op.The patient was confirmed to be unharmed.
 
Additional Manufacturer Narrative
The actual complaint product was not returned for evaluation.A review of the device history record is in-progress.All information reasonably known as of 28 sep 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint: (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Additional Manufacturer Narrative
The device history record for lot 20039944 was reviewed and the product was produced according to product specifications.Root cause could not be determined.All information reasonably known as of 16 nov 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ON-Q PAIN RELIEF SYSTEM, SOAKER CATHETER, NRFIT CONNECTOR, 270 ML
Common Device Name
ELASTOMERIC LFR
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT S. DE R.L. DE C.V.
ave noruega edificio d-1b
fraccionamiento rubio
tijuana b.c.
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key15508263
Report Number2026095-2022-00109
Device Sequence Number5171465
Product Code MEB
UDI-Device Identifier00193494450717
UDI-Public00193494450717
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K063530
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/08/2023
Device Model NumberPS2504-N
Device Catalogue NumberN/A
Device Lot Number20039944
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2022
Supplement Date Received by Manufacturer11/09/2022
Initial Report FDA Received Date09/29/2022
Supplement Report FDA Received Date11/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/23/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age45 YR
Patient SexMale
Patient Weight79 KG
-
-