Model Number 24686
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The stenosed target lesion was located in arteriovenous fistula.A 7.0 x80, 75cm gladiator elite balloon catheter was advanced for dilation.However, during the procedure, the balloon ruptured.The device was removed and the procedure was completed with another of the same device.No patient complications were reported and the patient was stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr: gladiator elite 7.0 x80, 75cm was received for analysis.The following attributes were examined: a visual examination identified that the balloon was not folded which indicates that the device was subjected to positive pressure.A balloon longitudinal tear was identified beginning approximately 3mm proximal of the proximal markerband and extending approximately 31mm distally across the balloon material.An examination of the balloon material and markerbands identified no issues which could potentially have contributed to this complaint.The rated burst pressure for this device is 20 atmospheres as per gladiator specification.A visual examination of the markerbands identified no issues which may potentially have contributed to the complaint incident.A visual and tactile examination found no kinks or damage to the shaft of the device.A visual examination identified no issues or damage to the tip of the device.This concludes the product analysis.
 
Event or Problem Description
It was reported that balloon rupture occurred.The stenosed target lesion was located in arteriovenous fistula.A 7.0 x80, 75cm gladiator elite balloon catheter was advanced for dilation.However, during the procedure, the balloon ruptured.The device was removed and the procedure was completed with another of the same device.No patient complications were reported and the patient was stable.It was further reported that the target lesion was 80% stenosed, severely calcified, and severely tortuous.
 
Additional Manufacturer Narrative
Device evaluated by mfr: gladiator elite 7.0 x80, 75cm was received for analysis.The following attributes were examined: a visual examination identified that the balloon was not folded which indicates that the device was subjected to positive pressure.A balloon longitudinal tear was identified beginning approximately 3mm proximal of the proximal markerband and extending approximately 31mm distally across the balloon material.An examination of the balloon material and markerbands identified no issues which could potentially have contributed to this complaint.The rated burst pressure for this device is 20 atmospheres as per gladiator specification.A visual examination of the markerbands identified no issues which may potentially have contributed to the complaint incident.A visual and tactile examination found no kinks or damage to the shaft of the device.A visual examination identified no issues or damage to the tip of the device.This concludes the product analysis.
 
Event or Problem Description
It was reported that balloon rupture occurred.The stenosed target lesion was located in arteriovenous fistula.A 7.0 x80, 75cm gladiator elite balloon catheter was advanced for dilation.However, during the procedure, the balloon ruptured.The device was removed and the procedure was completed with another of the same device.No patient complications were reported and the patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLADIATOR ELITE
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15509425
Report Number2124215-2022-38624
Device Sequence Number17652564
Product Code LIT
UDI-Device Identifier08714729809838
UDI-Public08714729809838
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K132810
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/26/2025
Device Model Number24686
Device Catalogue Number24686
Device Lot Number0028720221
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/14/2022
Supplement Date Received by Manufacturer11/03/2022
11/07/2022
Initial Report FDA Received Date09/30/2022
Supplement Report FDA Received Date11/08/2022
11/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/27/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age77 YR
Patient SexMale