• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ONKOS SURGICAL ELEOS; CEMENTED SEGMENTAL STEM, STRAIGHT, FLUTED, 17X120MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ONKOS SURGICAL ELEOS; CEMENTED SEGMENTAL STEM, STRAIGHT, FLUTED, 17X120MM Back to Search Results
Model Number HC-17120-03M
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 08/31/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient developed an alleged infection.The patient underwent a revision surgery on (b)(6) 2022 where the following eleos implants were explanted: tibial baseplate, cemented segmental stem, distal femur, cemented stem extension, modular collar locking ring, modular reduced resection collar, tibial poly spacer, tibial hinge component, and distal femur axial pin.No additional information regarding this adverse event has been reported.
 
Additional Manufacturer Narrative
The investigation is in process.The device will not be returned for evaluation.When the investigation is complete, a supplemental mdr will be submitted accordingly.Multiple mdrs were submitted for this event: #3013450937-2022-00274, #3013450937-2022-00275, #3013450937-2022-00276, #3013450937-2022-00277, #3013450937-2022-00278, #3013450937-2022-00279, #3013450937-2022-00281, #3013450937-2022-00282.The following mdrs were submitted for the same patient: #3013450937-2022-00087, #3013450937-2022-00088, #3013450937-2022-00089, #3013450937-2022-00090, #3013450937-2022-00091, #3013450937-2022-00092, #3013450937-2022-00093, #3013450937-2022-00094, #3013450937-2022-00095.
 
Additional Manufacturer Narrative
This report is being submitted to include additional information.The investigation is complete.The device was not returned for evaluation.The root cause of the alleged infection could not be determined.The device history records and sterilization batch records were reviewed and no issues during manufacturing or sterilization were identified that would have contributed to this complaint.If any additional information is obtained, a supplemental mdr will be filed accordingly.
 
Event or Problem Description
It was reported that the patient developed an alleged infection.The patient underwent a revision surgery on (b)(6) 2022 where the following eleos implants were explanted: tibial baseplate, cemented segmental stem, distal femur, cemented stem extension, modular collar locking ring, modular reduced resection collar, tibial poly spacer, tibial hinge component, and distal femur axial pin.No additional information regarding this adverse event has been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEOS
Common Device Name
CEMENTED SEGMENTAL STEM, STRAIGHT, FLUTED, 17X120MM
Manufacturer (Section D)
ONKOS SURGICAL
77 east halsey road
parsippany NJ 07054
Manufacturer (Section G)
PHILLIPS PRECISION INC
7 paul kohner pl
elmwood park NJ 07407
Manufacturer Contact
sara dailey
77 east halsey road
parsippany, NJ 07054
MDR Report Key15509560
Report Number3013450937-2022-00280
Device Sequence Number5069926
Product Code KRO
UDI-Device IdentifierB278HC1712003M0
UDI-PublicB278HC1712003M0
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K203815
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberHC-17120-03M
Device Catalogue NumberHC-17120-03M
Device Lot Number88744-001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/30/2022
Supplement Date Received by Manufacturer11/21/2022
Initial Report FDA Received Date09/30/2022
Supplement Report FDA Received Date11/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
P/N 25000007E, ELEOS DISTAL FEMUR; P/N 25001208E, ELEOS POLY SPACER; P/N 25002111E, ELEOS DISTAL FEMUR AXIAL PIN; P/N HR-30000-03M, ELEOS COLLAR LOCKING RING; P/N KSC01665E, ELEOS CEMENTED STEM EXTENSION; P/N SB-4044V-03M, ELEOS MODULAR COLLAR; P/N TB-2205E-01M, ELEOS TIBIAL BASEPLATE; P/N THSWMRS01M, ELEOS TIBIAL HINGE COMPONENT
Outcome Attributed to Adverse Event Hospitalization;
Patient Age66 YR
Patient SexMale
-
-