COOK INC NCIRCLE TIPLESS STONE EXTRACTOR; FFL DISLODGER, STONE, BASKET, URETERAL, METAL
|
Back to Search Results |
|
Catalog Number NTSE-045065-UDH |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/13/2022 |
Event Type
malfunction
|
Event Description
|
As reported, the basket wire of a ncircle tipless stone extractor was found broken when opening the package.A photo provided by the user appears to show multiple areas of damage on the basket coil.The procedure was completed by using another new device.No adverse effect to the patient reported.
|
|
Manufacturer Narrative
|
Customer occupation = unknown.Pma/510(k) #- exempt.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Manufacturer Narrative
|
Event description: as reported, the basket wire of a ncircle tipless stone extractor was found broken when opening the package.A photo provided by the user appears to show multiple areas of damage on the basket coil.The procedure was completed by using another new device.No adverse effect to the patient reported.Investigation ¿ evaluation.A visual inspection of the returned device was conducted.A document based investigation was also performed including a review of complaint history, device history record (dhr), manufacturing instructions, the instructions for use (ifu), and quality control procedures.One ncircle tipless stone extractor was returned in an open package.Inspection of the returned device noted: the basket assembly was separated from the handle.The basket sheath was damaged approximately 2mm from the flare.A review of the device history record found no non-conformances related to the reported failure mode.A review of complaint history records shows no other complaints associated with the complaint device lot.Because there were no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in the field.A review of relevant manufacturing and quality control documents was conducted.All extractors are inspected for damage and verified to assure the basket opens and closes properly.Cook has concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The instructions for use (ifu), provides the following information to the user related to the reported failure mode: precaution: do not use excessive force to manipulate this device.Damage to the device may occur.The returned device was found to be severely damaged.The cause for the damage is unknown.Excessive force may have been inadvertently applied to the device; however, no information is known regarding device handling, therefore the cause of the issue could not be conclusively determined.Per the quality engineering risk assessment, no further action is required.The appropriate personnel have been notified, and we will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
|
|