MAUDE Adverse Event Report: ABBOTT VASCULAR OMNILINK ELITE PERIPHERAL STENT SYSTEM; PERIPHERAL STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR OMNILINK ELITE PERIPHERAL STENT SYSTEM; PERIPHERAL STENT DELIVERY SYSTEM Back to Search Results
Model Number 1012625-19
Medical Device Problem Codes Difficult to Remove (1528); Difficult to Advance (2920); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was received.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat a heavily tortuous, moderately calcified right common iliac artery.The 9.0x19mm omnilink elite delivery system was noted to have met resistance with anatomy during advancement and removal of the device.Once removed it was noted that the stent dislodged on the delivery system.Not completely off but not well placed.Another same size ominilink was used to complete the procedure.There were no adverse patient effects and no clinically significant delay.No additional information was provided.
 
Additional Manufacturer Narrative
A visual inspection was performed on the returned device.The reported stent dislodgement was confirmed.The reported difficult to advance and difficult to remove could not be evaluated as the exact anatomical conditions encountered by the device used during the procedure could not be replicated in the test laboratory.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNILINK ELITE PERIPHERAL STENT SYSTEM
Common Device Name
PERIPHERAL STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15510531
Report Number2024168-2022-10156
Device Sequence Number5843849
Product Code NIO
UDI-Device Identifier08717648179327
UDI-Public08717648179327
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
P110043
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2022
Device Model Number1012625-19
Device Catalogue Number1012625-19
Device Lot Number9071141
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/08/2022
Supplement Date Received by Manufacturer10/06/2022
Initial Report FDA Received Date09/30/2022
Supplement Report FDA Received Date10/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/11/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age81 YR
Patient SexMale
Patient Weight76 KG
-
-