| Model Number |
63501 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/25/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Alleged deflation.
|
| |
|
Additional Manufacturer Narrative
|
|
Explant analysis showed silicone debris in the posterior valve channel, affection seal of the valve and causing a leak.
|
| |
|
Event or Problem Description
|
|
Deflation.
|
| |
|