| Model Number |
0502503010 |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It reported that there was blurry image.
|
| |
|
Additional Manufacturer Narrative
|
|
This scope was not received for evaluation at stryker endoscopy.This scope was received at henke for evaluation.Based on the henke service record attached, the reported failure "blurry" was confirmed.According to henke: scope evaluated as a level 3.Bent needle, dent in needle, distal tip/fiber damage, broken lenses in system, scratches on needle, residue on external optics.Probably root cause for this failure is: the complaint for ¿blurry¿ has been confirmed and is due to customer use and handling.The device manufacture date is not known.The reported failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It reported that there was blurry image.
|
| |
|