| Model Number |
8884720205 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that via a gastrostomy, when attempting to disconnect, the enteral nutrition set breaks the delivery line from the tube adapter.The device was replaced with another device with the same characteristics.There was no medical intervention required and no injury to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Additional Manufacturer Narrative
|
|
Since a lot number was not provided, the device history record review could not be performed.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.A sample was not received for the investigation.Because a sample was not returned, we were unable to perform a follow up investigation to include functional and visual evaluations to confirm the issue and root cause analysis.At this time, a corrective and preventive action is not deemed necessary.We will keep monitoring the process for any adverse trends that require immediate attention.This complaint will be used for qa tracking and trending purposes.
|
| |
|