| Model Number |
010911L |
| Medical Device Problem Codes |
Entrapment of Device (1212); Failure to Form Staple (2579); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Tissue Injury (4559)
|
| Date of Event |
09/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic donor nephrectomy, while separating the renal artery and renal vein, the tissue did not come off from the cartridge when the jaw was opened after firing; it did not come off when an attempt was made to remove it using forceps, the jaw of device was fixated to tissue. an additional separation was attempted using powered stapler from the planned resection position during renal vein separation, however, the already bitten device did not work out well due to interference. the staple that was left in the cartridge fork is not a clean b-type. to resolve the issue and complete the case, emergency transition from laparoscopic surgery to laparotomy was done and the incision was expanded wider than expected, exceeded 1 inch (2.54 cm), the blood vessel was sutured with a needle suture, then cut with a scalpel then collect the tissue and the cartridge, and the surgery time was extended by 30 minutes or more.
|
| |
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: 010911l - 010911l multifire endo ta* 30-2.5 dlu x6, lot# p0c1213my.010901 - 010901 multifire endo ta 30-2.5 x3, lot# p0h0188y.Sigphandle - sig power sigphandle handle, lot# unknown.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted the loading unit displayed a full complement of staples and the interlock was engaged.Functionally, the instrument interlock was reset and the instrument was applied with no abnormalities.It was reported that the laparoscopic surgery was required to be converted to an open surgery.Operating (or) time was extended by >30 minutes resulting from product failure.The staples did not form as expected and the jaws of the device remain closed on tissue.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur when the loading unit was improperly loaded.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: when inserting a loading unit into the multifire endo ta stapler, ensure that the loading unit is fully inserted as signified by an audible click and alignment of the black lines on the loading unit and the anvil.Failure to properly insert the loading unit into the stapler may result in malformed staples.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|