| Model Number |
A3101 |
| Medical Device Problem Code |
Mechanics Altered (2984)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
A facility reported that the mayfield composite series base unit (a3101) was wobbling during a "craniotomy for lesion" procedure, the "teeth" or starbursts was not gripping in.There was a minimal delay in surgery to stabilize the unit.The patient had already been pinned in the clamp, so they continued with the case.There was no patient injury or death reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The mayfield composite series base unit (a3101) was returned for evaluation: device history record (dhr)- the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - investigation showed that the two halves of the swivel adaptor do not close, and the paint had bubbled.The estimated life expectancy for mayfield products is 7 years in consideration of all the variables that can influence the life and function of these devices.Root cause - complaint confirmed via inspection of the unit.Probable root cause is routine use and wear.Additionally the paint had bubbled on the base unit and as a result the unit was replaced for a new a3101.Corrective and preventative action (capa) related to the powder coat peeling on the a3101 base units is in process; therefore, this complaint is being closed.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|