Model Number PM2272 |
Device Problems
Premature Discharge of Battery (1057); Pacing Asynchronously (1441)
|
Patient Problems
Arrhythmia (1721); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Further information was requested but not received.
|
|
Event Description
|
During an in clinic follow-up, the device longevity was shorter than expected.Premature depletion of the battery was suspected.No intervention has been performed at this time.The patient was stable.
|
|
Event Description
|
Further review of the records by abbott technical support indicate that the increased battery depletion appears normal and was due to the device often having a high ventricular output and no intervention was required to resolve the event.Additionally, the event of pacemaker mediated tachycardia (pmt) was observed on the device.
|
|