Catalog Number UNK REAMER
Device Problems Corroded (1131); Contamination /Decontamination Problem (2895)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/27/2022
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).The device catalog number is unknown; therefore, udi is unavailable.Investigation summary the device associated with this report was not returned for analysis.Review of the photographic evidence confirmed the complaint.Foreign substance or stains can be observed on the instrument.It is reasonable to conclude that the substance is corrosion.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The hospital received dirty instruments.(srom instruments).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNK REAMER
Type of Device
REAMERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key15513370
MDR Text Key301503328
Report Number1818910-2022-19216
Device Sequence Number1
Product Code HTO
Combination Product (y/n)N
Reporter Country CodeIT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK REAMER
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/22/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-