BOSTON SCIENTIFIC CORPORATION DIREXION HI-FLO FATHOM-16 SYSTEM; CATHETER, CONTINUOUS FLUSH Back to Search Results
Model Number 84591
Medical Device Problem Codes Entrapment of Device (1212); Stretched (1601)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluated by mfr.: the direxion microcatheter was returned for analysis.The device was stuck in the angiographic catheter that was used for the procedure.It was noticed that the shaft was fractured located 46.5cm and 55cm from the hub.The returned angio catheter was a 5fr.035 internal diameter.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.The device was confirmed for being stuck in the returned angiographic catheter, as well as shaft fractures.
 
Event or Problem Description
Reportable based on device analysis completed on 08-sep-2022.It was reported that the catheter got stuck and stretched.A 130cm direxion hi-flo catheter-infusion was selected for use.However, the device appeared to get stuck inside the base of the catheter and stretched.It took some time to get the catheter positioned into the original location.No patient complications were reported.However, returned device analysis revealed shaft was fractured.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIREXION HI-FLO FATHOM-16 SYSTEM
Common Device Name
CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15513547
Report Number2124215-2022-37882
Device Sequence Number17453664
Product Code KRA
UDI-Device Identifier08714729839712
UDI-Public08714729839712
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K142259
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/21/2024
Device Model Number84591
Device Catalogue Number84591
Device Lot Number0028721418
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/08/2022
Initial Report FDA Received Date09/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1