| Model Number |
84591 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Stretched (1601)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: the direxion microcatheter was returned for analysis.The device was stuck in the angiographic catheter that was used for the procedure.It was noticed that the shaft was fractured located 46.5cm and 55cm from the hub.The returned angio catheter was a 5fr.035 internal diameter.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.The device was confirmed for being stuck in the returned angiographic catheter, as well as shaft fractures.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 08-sep-2022.It was reported that the catheter got stuck and stretched.A 130cm direxion hi-flo catheter-infusion was selected for use.However, the device appeared to get stuck inside the base of the catheter and stretched.It took some time to get the catheter positioned into the original location.No patient complications were reported.However, returned device analysis revealed shaft was fractured.
|
| |
|