AIZU OLYMPUS CO., LTD. GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-XP150N
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
During inspection and testing, the reported issue (reduced angulation) was confirmed.It was observed the bending angel in up/down directions was out of specification due to wear of the angulation wire.In addition, service found the objective lens was scratched, the adhesive around the light guide lens was worn, the distal end was dirty, the adhesive on the bending section cover was detached, the connecting tube was wrinkled, and the image had shadows due to damaged ccd (charge-coupled device) unit.The forceps could not be inserted smoothly, and the channel cleaning brush could not be inserted smoothly due to damaged instrument channel.The investigation is ongoing; therefore, a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Event or Problem Description
The customer reported there was play in the angulation knob and the bending angle was reduced.There was no patient or user injury reported due to the event.The subject device was returned to an olympus service center for evaluation.Upon inspection and testing of the returned device, foreign material was observed in the nozzle.This report is being submitted for the malfunction found during device evaluation (foreign material in nozzle).Additional details have been requested regarding the reported issue.At this time, no additional information has been provided.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information that was received from the user facility regarding the reported issue.
 
Event or Problem Description
Additional information was received from the user facility regarding the reported event (play in the angulation knob and reduced bending angle): the reported issue was observed during routine/ general inspection, prior to an unspecified procedure.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified; however, it was confirmed that foreign matter was attached to the nozzle.It was confirmed that there was no deformation of the air/water nozzle and there were no obvious deviations in reprocessing.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GASTROINTESTINAL VIDEOSCOPE
Common Device Name
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer Contact
masaharu hirose
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8-520
JA   965-8520
426422891
MDR Report Key15514240
Report Number9610595-2022-02427
Device Sequence Number12242404
Product Code FDS
UDI-Device Identifier04953170229442
UDI-Public04953170229442
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K051645
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberGIF-XP150N
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2022
Supplement Date Received by Manufacturer01/17/2023
03/08/2023
Initial Report FDA Received Date09/30/2022
Supplement Report FDA Received Date01/20/2023
04/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/03/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1