• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ PANOPTIX TORIC TRIFOCAL IOL; LENS, MULTIFOCAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ PANOPTIX TORIC TRIFOCAL IOL; LENS, MULTIFOCAL Back to Search Results
Model Number TFNT30
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Visual Impairment (2138)
Date of Event 05/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A sample device was not returned for analysis.Product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported following an intraocular lens implant procedure, the patient experienced refractive outcome problem.The lens was later exchanged in a secondary procedure with a mono focal lens.Additional information has been requested.
 
Additional Manufacturer Narrative
The account indicated the use of a qualified cartridge, handpiece and viscoelastic.The product investigation could not identify a root cause for the reported complaint.Each lens is subjected to a 100% assessment of the power and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Information was provided that the multifocal company (1.50cyl), 22.5 diopter (add power +2.2/+3.2) lens was replaced with a non-company, 23.5 diopter, monofocal lens model.Due to the exchange of a toric-multifocal lens to a non-toric-monofocal lens, this may suggest that the replacement lens model was an improved choice for the patient's vision needs.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Additional information was received stating the implantation of new iol took place on same day of previous lens explant.Explanted lens was not available for return and investigation since lens was cut during explantation.Physician reported that it is not a problem of calculation of the company image guided system because the postoperative refraction was good but the visual acuity was bad.It is either a problem of tolerance of the multifocality or a problem of the implant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ PANOPTIX TORIC TRIFOCAL IOL
Common Device Name
LENS, MULTIFOCAL
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15517244
Report Number1119421-2022-02034
Device Sequence Number2533226
Product Code MFK
UDI-Device Identifier00380652390026
UDI-Public00380652390026
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
P040020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/07/2026
Device Model NumberTFNT30
Device Catalogue NumberTFNT30.225
Device Lot Number15207746
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/31/2022
Supplement Date Received by Manufacturer11/02/2022
Initial Report FDA Received Date09/30/2022
Supplement Report FDA Received Date12/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNSPECIFIED MONOFOCAL LENS; VERION DIGITAL MARKER M
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-