• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA REVERSE SHOULDER SYSTEM 04.01.0116 HUMERAL REVERSE HC LINER Ø32/+0MM; SHOULDER HUMERAL REVERSE HC LINER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA REVERSE SHOULDER SYSTEM 04.01.0116 HUMERAL REVERSE HC LINER Ø32/+0MM; SHOULDER HUMERAL REVERSE HC LINER Back to Search Results
Model Number 04.01.0116
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Pain (1994)
Date of Event 09/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 12-sep-2022.Lot 179975: (b)(4) items manufactured and released on 29-mar-2018.Expiration date: 2023-03-19.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with another similar reported case during the period of review.
 
Event or Problem Description
At about 1 month after the primary surgery, the patient came in the emergency room reporting pain due to dislocation and the cause is unknown.The emergency room doctor reduced the patient and put her in a shoulder sling.When the patient went to a visit with the operating surgeon, he was planning on cementing a new stem, but after opening the patient he decided that the stem was solid and was not going to subside.The surgeon revised liner from a +0 to +6.The surgery was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVERSE SHOULDER SYSTEM 04.01.0116 HUMERAL REVERSE HC LINER Ø32/+0MM
Common Device Name
SHOULDER HUMERAL REVERSE HC LINER
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key15518430
Report Number3005180920-2022-00737
Device Sequence Number14959722
Product Code PHX
UDI-Device Identifier07630040706230
UDI-Public07630040706230
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K170452
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/19/2023
Device Model Number04.01.0116
Device Catalogue Number04.01.0116
Device Lot Number179975
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2022
Initial Report FDA Received Date09/30/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/29/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-