| Model Number |
4068310 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer alleged the light head was dropped.The bio med went to secure the light head to the spring arm and it fell to the floor.No harm or impact to a surgical procedure was reported.This report was filed in our complaint handling system as complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Inspection of the device is pending.Further investigation should confirm the root cause.Investigation results will be provided within a final report if new information will become available.
|
| |
|
Event or Problem Description
|
|
The customer alleged the light head was dropped.The bio med went to secure the light head to the spring arm and it fell to the floor.No harm or impact to a surgical procedure was reported.This report was filed in our complaint handling system as complaint #(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
During follow-up with the customer it was identified the light head fell during the maintenance procedure carried out by the hospital biomed.For maintenance it is required to remove the light head from the support arm system.During this, the light head fell based on a handling failure of the technician.There was no impact to a surgical procedure and no identified device malfunction.Mthe defect light head will be replaced.Based on this, not further actions are required.
|
| |
|