• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC SIMPLE PNEUMOTHORAX ASPIRATION ACCESSORY SET; JOL CATHETER AND TIP, SUCTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC SIMPLE PNEUMOTHORAX ASPIRATION ACCESSORY SET; JOL CATHETER AND TIP, SUCTION Back to Search Results
Model Number N/A
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/31/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.Initial reporter occupation: clinical ops specialist.Pma/510(k) #: exempt.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
Three unused chest drain valves from simple pneumothorax aspiration accessory sets were returned to cook.It was reported that the valve leaflets were adhered to themselves preventing air flow.
 
Additional Manufacturer Narrative
Investigation-evaluation: (b)(6) informed cook of an issue where the chest drain valves within simple pneumothorax aspiration accessory sets (c-casp-a-ford) were adhered to themselves and would not allow air to flow through them.To aid in investigation, the customer returned 13 unused/prior to use devices.The devices were returned on 20sep2022, and complaints were opened to capture the malfunctioning devices by lot number.This complaint captures the three unused devices from lot 14762352.The issue was observed in cook's biohazard lab and there was no patient involvement.Reviews of documentation including the complaint history, device history record, drawing, quality control, specifications, trends and device history record of the device, as well as a visual inspection, functional test, and inspection of unused product, were conducted during the investigation.The device master record (dmr) was reviewed and device drawings, quality control procedures, and specifications were identified.Sufficient controls are in place to detect to prevent the release of nonconforming product.The user facility returned thirteen devices to aid in investigation of this issue (previous incidents captured under 1820334-2022-01495, 1820334-2022-01496, 1820334-2022-01498, 1820334-2022-01497).Two opened, prior to use devices were returned as well as ten sealed and one in an open inner pouch.A syringe was used to test the devices.Air did not pass through and there was no movement of the inner valve.Water was then pushed through the syringe, and water passed through.However, the valve appeared to stay stuck near the corners and would not open entirely.This was observed in all thirteen returned devices.The returned devices are captured in unique complaints by lot number (1820334-2022-01497, 1820334-2022-01564, 1820334-2022-01566, 1820334-2022-01567, 1820334-2022-01568, 1820334-2022-01569, 1820334-2022-01570).This report captures the three unused devices from lot 14762352.A review of the device history record (dhr) showed no related nonconformances.A database search revealed no other complaints under this lot number at the time of this investigation.An expanded search of complaint history has shown no incidents of the inner valve adhering to itself/unable to pass air other than those reported by the same user facility.A review of nonconformances on this rpn and the chest drain valve sold individually (c-cdv-1) from 01oct2020 through 01jul2022 did not show relevant nonconformances.The chest drain valve component is supplied to cook by an external supplier.There were no abnormal processing conditions nor manufacturing related issues identified by the supplier or sub-supplier.The sub-supplier identified transportation/storage as a possible cause of failure.The information provided upon review of the dmr, dhr, returned devices, and supplier investigation suggests that the devices were manufactured to specification, and that there are no nonconforming devices in house or out in the field.The complaint was confirmed based on customer testimony and investigation of the returned devices.The user was unable to provide the storage conditions of the failed devices, however the user did indicate that these devices are used during air ambulance transport cases.Cook has concluded the most likely cause of failure is related to storage conditions.The risk assessment for this failure mode was reviewed, and no action is required.The appropriate personnel were notified, and cook will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding patient and/or event details has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIMPLE PNEUMOTHORAX ASPIRATION ACCESSORY SET
Common Device Name
JOL CATHETER AND TIP, SUCTION
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key15518628
Report Number1820334-2022-01566
Device Sequence Number2674446
Product Code JOL
UDI-Device Identifier00827002052970
UDI-Public(01)00827002052970(17)240601(10)14762352
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/01/2024
Device Model NumberN/A
Device Catalogue NumberC-CASP-A-FORD
Device Lot Number14762352
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/06/2022
Supplement Date Received by Manufacturer04/02/2023
Initial Report FDA Received Date09/30/2022
Supplement Report FDA Received Date04/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-