Model Number 1804300-28
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A visual and dimensional inspection was performed on the returned device which identified the shaft was stretched.The reported failure to advance could not be evaluated as the exact anatomical conditions encountered by the device used during the procedure could not be replicated in the test laboratory.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot specific quality issue.The investigation determined the reported failure to advance appears to be related to circumstances of the procedure, as it is likely that the device interacted with the mildly tortuous lesion during advancement, as resistance was reported, which contributed to the reported failure to advance and observed material deformation (flared strut).Manipulation of the device during retraction likely contributed to the observed stretched inner/outer member.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Event or Problem Description
It was reported that the procedure was to treat a lesion in the mid left anterior descending (lad) artery with mild tortuosity.The 3.0x28mm xience skypoint stent delivery system (sds) met resistance during advancement and failed to cross the lesion due to the anatomy.There was no resistance during removal.There were no other device issues such as shaft separation or stent dislodgement.There was no adverse patient effect.There was no clinically significant delay in the procedure.No additional information was provided.Returned device analysis identified a stretched shaft.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15519776
Report Number2024168-2022-10182
Device Sequence Number12269167
Product Code NIQ
UDI-Device Identifier08717648233265
UDI-Public08717648233265
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/20/2024
Device Model Number1804300-28
Device Catalogue Number1804300-28
Device Lot Number107144A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/27/2022
Initial Report FDA Received Date09/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-