MAUDE Adverse Event Report: ABBOTT VASCULAR RX HERCULINK ELITE STENT SYSTEM; STENT, RENAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR RX HERCULINK ELITE STENT SYSTEM; STENT, RENAL Back to Search Results
Model Number 1011487-15
Medical Device Problem Codes Difficult to Insert (1316); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A visual inspection was performed on the returned device.The reported difficult to insert was not tested due to the condition of the returned device; however, it was noted that the stent was dislocated on the balloon.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents.The investigation was unable to determine a conclusive cause for the reported difficult to insert; however, the observed dislocated stent was likely due to the reported difficult to insert.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that the procedure was to treat a mildly calcified, mildly tortuous, 90% stenosed lesion in the left renal (lr) artery.A 4x15mm herculink elite balloon expandable stent system (bess) was unable to be advanced into an unspecified 6f guiding catheter (gc) after several attempts.There was no adverse patient effect and no clinically significant delay in the procedure.Another same size bess was used to successfully complete the procedure.Returned device analysis noted that the stent moved, was dislocated on the balloon.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX HERCULINK ELITE STENT SYSTEM
Common Device Name
STENT, RENAL
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15520480
Report Number2024168-2022-10185
Device Sequence Number2690360
Product Code NIN
UDI-Device Identifier08717648077838
UDI-Public08717648077838
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P110001
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2022
Device Model Number1011487-15
Device Catalogue Number1011487-15
Device Lot Number0010761
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/25/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/21/2022
Initial Report FDA Received Date09/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/07/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-