• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. BIVONA FLEXTEND PEDIATRIC NEONATAL TRACHEOSTROMY; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. BIVONA FLEXTEND PEDIATRIC NEONATAL TRACHEOSTROMY; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 670160
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A product sample was received and is awaiting evaluation and investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event or Problem Description
It was reported that prior to insertion with a new cuff, cuff balloon silicone lifted away from where syringe is inserted into.Silicone will not go back into place.No injury was reported; no medical intervention.
 
Additional Manufacturer Narrative
Other text: h6: event problem and evaluation codes: updated.H10: a device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual inspection found the valve improperly place from pilot balloon was found; complaint was confirmed.No other analysis was performed.The root cause of the reported issue was found to be assembly fault - not fully seated the valve or too far in with potential cause of not follow the instructions.Customer complaint notification was made to production personnel as awareness notification for the reported failure mode, by quality engineer on 31/jan/2023.This remediation mdr was generated under protocol b10010579, as a result of warning letter cms# 617147.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA FLEXTEND PEDIATRIC NEONATAL TRACHEOSTROMY
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
NULL
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15521647
Report Number3012307300-2022-20574
Device Sequence Number6079212
Product Code JOH
UDI-Device Identifier15021312006032
UDI-Public15021312006032
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K944178
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number670160
Device Catalogue Number670160
Device Lot Number4218780
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/08/2022
Supplement Date Received by Manufacturer02/13/2023
Initial Report FDA Received Date09/30/2022
Supplement Report FDA Received Date03/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/05/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-