| Model Number |
BX70071D |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/23/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 01/14/2020, a user facility representative reported a complaint that "an administration set model n/a lot # also n/a is leaking right below the drip chamber.This happened three separate times and one of the sets is being discarded." a patient was involved but was not harmed or injured.The device operator was a registered nurse.The medication being infused was saline and iron sucrose.The contract manufacturer of the affected device is (b)(4) medical co.Ltd.This complaint was determined as reportable during a retrospective review performed in (b)(6) 2022.This mdr is being submitted as part of remediation actions.
|
| |
|
Additional Manufacturer Narrative
|
|
The service provider provided the evaluation report on 01/29/2021.Three used administration sets (bx70071d lot# 1811011) were returned from the customer and tested.Leaking below the drip chamber was observed.The reported issue was confirmed.
|
| |
|
Event or Problem Description
|
|
This is a follow-up for the initially filed mdr (3006575795-2022-00041).
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: remove "this complaint was determined as reportable during a retrospective review performed in september 2022.This mdr is being submitted as part of remediation actions." in b5 of the initial mdr report #3006575795-2022-00041.Additional information: mdr report #3006575795-2022-00041 was submitted september 30, 2022, for complaint (b)(4).Mdr report #3006575795-2022-00041 was submitted to document an individual incident of the three (3) submitted under a single mdr report #3006575795-2021-00002, complaint #(b)(4).
|
| |
|