BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 382537
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/12/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event is unknown; awareness date has been used for this field.Initial reporter address information was not able to be obtained, therefore, (b)(6) was used as a place holder.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported while using bd insyte¿ autoguard¿ bc shielded iv catheter the needle would not retract.There was no report of patient impact.The following information was provided by the initial reporter: the button will not retract the needle.
 
Additional Manufacturer Narrative
H6: investigation summary since no samples displaying the condition reported were available for examination, we were unable to fully investigate this incident.The dhr for lot 2013579 has been reviewed.Subassembly lot 2013579 material 7000bcs37 was built on afa line 9 from (b)(6)2022 through (b)(6)2022.Final lot was packaged on pkg line 11 from (b)(6)2022 through (b)(6)2022 and pkg line 8 from (b)(6)2022 through (b)(6)2022 for a total quantity of 209,210 units.Qn#(b)(4).Was initiated for damage grips during the built of lot number, which could potentially be related to this complaint.No process deviations were found.
 
Event or Problem Description
It was reported while using bd insyte¿ autoguard¿ bc shielded iv catheter the needle would not retract.There was no report of patient impact.The following information was provided by the initial reporter: the button will not retract the needle.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer (Section G)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15522772
Report Number1710034-2022-00569
Device Sequence Number6029408
Product Code FOZ
UDI-Device Identifier00382903825370
UDI-Public00382903825370
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K110443
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2024
Device Catalogue Number382537
Device Lot Number2013579
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2022
Supplement Date Received by Manufacturer11/01/2022
Initial Report FDA Received Date10/01/2022
Supplement Report FDA Received Date11/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-