NEUROPACE, INC. NEUROPACE RNS SYSTEM Back to Search Results
Model Number DL-330-10-K
Medical Device Problem Codes Signal Artifact/Noise (1036); Impedance Problem (2950)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Neuropace is pending return of the explanted lead.
 
Event or Problem Description
On (b)(6) 2022, a neuropace depth lead was observed to have insufficient charge on all electrode contacts(4), signal artifact, and abnormal impedances, indicative of a lead break.This was confirmed by neuropace after review of the intracranial eeg recordings from that lead.Neuropace confirmed that the data was indicative of a lead break.The lead was explanted and replaced on (b)(6) 2022, and functioned normally.The patient is reported as doing well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROPACE RNS SYSTEM
Common Device Name
NEUROPACE RNS SYSTEM
Manufacturer (Section D)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer (Section G)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer Contact
ramona gonis
455 n. bernardo ave.
mountain view, CA 94043
6502382788
MDR Report Key15524040
Report Number3004426659-2022-00036
Device Sequence Number19501743
Product Code PFN
UDI-Device Identifier00855547005045
UDI-Public010085554700504517220411
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P100026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberDL-330-10-K
Device Catalogue Number1007605
Device Lot Number27875-1-1-1
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/06/2022
Initial Report FDA Received Date10/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Hospitalization;
Patient Age38 YR
Patient SexMale
-
-