• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIDENT 0 DEG INSERT 40MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIDENT 0 DEG INSERT 40MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 623-00-40E
Medical Device Problem Codes Degraded (1153); Device Dislodged or Dislocated (2923); Naturally Worn (2988)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Joint Dislocation (2374); Ambulation or Postural Difficulties (2544)
Date of Event 09/07/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery was performed due to head dislocation and stem neck shaving.A black discoloration of the bone and soft tissue within the joint were revealed at the revision.There was no bone fragility or soft tissue loss.The cup continued to be used and the insert was replaced due to age-related wear.
 
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned to the manufacturer.
 
Additional Manufacturer Narrative
Reported event: an event regarding wear involving a trident liner was reported.The event was confirmed via evaluation of the returned device.Method & results: -product evaluation and results: visual inspection of the device indicated that the outer surface of the liner which articulates with the shell is worn.Damage is consistent with delamination and material loss.The yellow discoloration observed on the liner is consistent with the absorption of synovial fluid by the device.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to disassociation of the head from stem.Liner wear was also reported.Visual inspection of the device indicated that the outer surface of the liner which articulates with the shell is worn.Damage is consistent with delamination and material loss.The yellow discoloration observed on the tibial insert is consistent with the absorption of synovial fluid by the device.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Revision surgery was performed due to head dislocation and stem neck shaving.A black discoloration of the bone and soft tissue within the joint were revealed at the revision.There was no bone fragility or soft tissue loss.The cup continued to be used and the insert was replaced due to age-related wear.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT 0 DEG INSERT 40MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
arokiya raj
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15525137
Report Number0002249697-2022-01421
Device Sequence Number6028956
Product Code LPH
UDI-Device Identifier07613327025323
UDI-Public07613327025323
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K062419
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2015
Device Model Number623-00-40E
Device Catalogue Number623-00-40E
Device Lot NumberMJL27Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/07/2022
Supplement Date Received by Manufacturer12/01/2022
Initial Report FDA Received Date10/01/2022
Supplement Report FDA Received Date12/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/23/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-